A proper sequence of dynamic alignment in transtibial prosthesis: insight through socket reaction moments

Jul 18, 2023

Scientific Reports volume 13, Article number: 458 (2023) Cite this article

2926 Accesses

8 Altmetric

Metrics details

Dynamic alignment in prosthetic fitting is important because it affects the user's stability, kinematics, and kinetics such as socket reaction moments. It is performed by tuning the spatial relationship between the transtibial prosthetic socket and the foot following sequential observational gait analysis in the three anatomical planes. However, the order of planes in which the adjustment should be performed is still unclear. To investigate the appropriate sequence of dynamic alignment adjustment, ten participants with transtibial amputation were asked to walk in different alignment conditions (flexion, extension, adduction, abduction; lateral, medial, anterior, and posterior translation of the socket, and plantarflexion, dorsiflexion, inversion, and eversion of the foot) to measure socket reaction moments in the out-of-planes (e.g., the effect of sagittal alignment on the coronal moment). A significant difference was found only among socket posterior translation, socket flexion, and baseline alignment in the coronal moment (P = 0.02). The results of the current and previous studies suggest that moments in the coronal plane are affected by alignment changes in all three planes, whereas moments in the sagittal plane are affected only by sagittal alignment changes. It is suggested that the procedure of alignment adjustments should be finalized in the coronal plane.

The spatial relationship between the prosthetic socket and the prosthetic foot of a transtibial prosthesis is defined as "prosthetic alignment"1. Prosthetic alignment influences stability, comfort, spatiotemporal, kinetic, and kinematic parameters during gait in individuals with transtibial amputation2,3,4,5,6,7,8,9,10,11. Prosthetic alignment consists of angular and translational adjustments of the components in the prosthetic device. Angular adjustments include flexion, extension, adduction, and abduction of the socket, as well as dorsiflexion, plantarflexion, inversion, eversion, and internal/external rotation (toe-in/out) of the prosthetic foot. Translational adjustments involve anterior, posterior, medial, and lateral translation of the socket relative to the foot1. Traditionally, prosthetic alignment is tuned in the following sequence: bench alignment, static alignment, and dynamic alignment12.

Bench alignment is the alignment set on a workbench before fitting the prosthesis. It has been recommended as follows: the socket is flexed and adducted approximately five degrees, and the vertical line from the antero-posterior center of the socket at the mid-patellar tendon level falls anterior to the heel breast12,13 or 18–65 mm anterior to the center of the foot14 in the sagittal plane. In the coronal plane, it is recommended that the vertical line from the medio-lateral center of the socket falls 10–30 mm lateral to the center of the heel12,13. Pre-bench alignment assessment procedures have also been reported such as the vertical alignment axis method (VAA)15,16,17 and anatomically-based alignment (ABA)14,16,17. Nowadays, endoskeletal prostheses are predominant in the market and manufacturers have guidelines regarding bench alignment for each prosthetic foot.

Static alignment is performed while an individual with transtibial amputation is standing upright in their prosthesis. Prosthetists check the levelness of the pelvis to assess the height of the prosthesis, and examine whether the prosthetic foot is placed flat on the ground. The user is asked to comment on the comfort level and stability during fitting and standing with the prosthesis. When the position of the foot is inappropriate and/or the posture/balance is poor, the alignment needs to be adjusted13. Previous studies have suggested that the vertical component of the ground reaction force18,19 or the use of anteroposterior radiographs20 may be useful for establishing static alignment appropriately.

After the static alignment has been completed, the dynamic alignment is evaluated and adjusted based primarily on observational gait analysis. Visual gait observation and perception of users with prostheses are commonly used in clinical practice to guide dynamic alignment. The prosthetist first considers the user's physical characteristics such as muscular strength, residual limb length, walking confidence and activity level. This is established using the observations by the prosthetists as well as feedback from the user of the prosthesis12,13. During the walking trials, prosthetists observe the gait of the individual with transtibial amputation to identify any kind of deviations in the sagittal, coronal and transverse planes. It should be noted that the gait deviations are generally considered to be linked with malalignment in the same plane as described in textbooks used for clinical education1,12,13. Prosthetists also ask the user for their comfort level when walking with the prosthesis. When a gait deviation is identified to be associated with malalignment, adjustments are made to minimize the deviation and/or discomfort, and then walking trials are performed again. This process is iterated until both the prosthetist and the user are satisfied21.

Previous studies have indicated that prosthetists’ judgement about the dynamic alignment may not be completely reliable and accurate. For example, feedback from prosthetic users such as their perception (e.g., comfort/complaint) about their prosthesis during walking is primarily expressed verbally, and the clinicians need to carefully interpret it and connect to potential contributing factors. It has also been reported that the perception of prosthetic users is not always accurate and does not necessarily reflect the prosthetic alignment changes22,23. Observation of gait deviation is focused on kinematic and temporo-spatial parameters. Though some joint angles were reported to be affected by alignment changes (e.g., increased internal rotation of prosthetic feet may increase maximum knee flexion angle)24, effects of alignment changes on kinematic parameters may not be predictable25. Similarly, temporo-spatial parameters may be influenced by alignment changes, but might not serve as a good predictor of prosthetic alignment changes25. These findings are consistent with a report by Zahedi et al., which showed that observational gait analysis of prosthetic alignment changes by prosthetists may not be reproducible26.

In contrast, kinetic parameters may be useful to evaluate deviations in prosthetic alignment. Ground reaction forces have been reported to be affected by prosthetic alignment changes during walking5,7,10. Moment of force measured with load cells displayed a strong correlation with intra-socket pressure10. Socket reaction moment, or external moment of force measured with an embedded load cell in prostheses, has been reported to be a good predictor to alignment changes of transtibial prostheses27,28,29. Chen et al. reported that the use of socket reaction moment would lead to a similar prosthetic alignment to conventional methods based on observation of prosthetists and feedback of the user although it resulted in slightly higher varus moments30.

Dynamic alignment needs to be addressed across all three anatomical planes (sagittal, coronal, and transverse), however, an agreement about their sequence and/or priority for quick and accurate prosthetic alignment has not been consistently reached in rigorous peer-reviewed research. Several authors have suggested a specific sequence for adjustments across each plane, but there has been little agreement of this sequence. It was recommended by Radcliffe et al. that dynamic alignment should be first tuned in the coronal plane, followed by the sagittal plane, because stability in the sagittal plane is crucial and should not be achieved until adjustments in the coronal plane have been completed12,13. It has also been reported that linear (or translational) adjustment should be performed first, followed by tilt (or angular) adjustment31. However, it is still unclear whether these sequences (e.g., first in the coronal, then sagittal plane, or translation first, then angulation) are reasonable. It has been reported that transtibial alignment changes in the sagittal plane significantly affect the moments in the coronal plane, however, alignment changes in the coronal plane did not reciprocally influence the sagittal plane moments32. Based on these findings, Kobayashi et al.32 suggested that dynamic alignment should be conducted first in the sagittal plane followed by the coronal plane. Thus, the socket reaction moment may be useful to determine the order of adjustment in the dynamic alignment. Furthermore, it should be noted that the conclusion of their study appears inconsistent with Radcliffe et al.'s recommendation. Also, their study32 did not clarify the differences in the effects between translational and angular alignment changes on moments when the displacement of the foot from the socket is equal (d in Fig. 1)23,33. The difference of effects in out-of-plane moments between angular changes and translational changes should be investigated with equal displacement in order to consider the order of angular and translational adjustments. Toe-in/toe-out angles of prosthetic feet have also been reported to significantly affect moments in the out-of-plane (e.g., in the coronal plane)29. It also should be clarified whether angulation changes of prosthetic feet in the sagittal or coronal plane (i.e., dorsiflexion/plantarflexion, or inversion/eversion) influence moments in the out-of-planes. Examining the effect of alignment changes of the socket or the foot on the socket reaction moments in out-of-planes could reveal the appropriate sequence of dynamic alignment adjustment in the transtibial prosthesis.

Equal displacement (d) of the prosthetic foot in angular changes and translational changes.

The aim of this study was to investigate the effects of alignment changes of transtibial prostheses on the out-of-plane socket reaction moments in order to seek the appropriate sequence of dynamic alignment adjustment. We hypothesized that the sequence of planes in dynamic alignment could be determined based on the out-of-plane effects of the alignment changes on the socket reaction moments29,32.

Ten participants (nine males and one female) were recruited who were also the same participants group from a previous study33 (Table 1). Their mean (SD: standard deviation) age, height, and body mass were as follows: 51.2 (13.5) years old, 170 (7.8) cm, and 67.7 (8.5) kg. The inclusion criteria were having a unilateral transtibial amputation and being a community ambulator without walking aids. The exclusion criteria were having any orthopedic and/or neurological disorders and being under the age of twenty. This study was approved by the Internal Review Board of the Graduate School of Comprehensive Rehabilitation, Osaka Prefecture University (Approval number: 2018–101) and conducted according to the Declaration of Helsinki guidelines. Written informed consent was obtained from all participants.

An instrumented prosthetic pyramid (Europa, Orthocare Innovations LLC, Edmonds,WA, USA) was used to measure the magnitude and timing of socket reaction moments30,34,35. A three-dimensional motion capture system (Vicon, Vicon Motion Systems Ltd., UK) with twelve infrared cameras and two force plates (AMTI, USA) was used with a Plug-in-Gait marker set36 to measure walking parameters including speed. The sampling rate was set at 100 Hz for the Europa and the motion capture system, and 1000 Hz for the force plates.

A clamp adapter, a slide adapter, a pylon, a prosthetic foot (LP Vari-Flex, Ossur HF, Reykjavik, Iceland), and the instrumented prosthetic pyramid were used to build an experimental prosthesis using the participant's own socket (Fig. 2). ID2 used PTB socket with a cuff suspension strap, whereas other participants used TSB sockets with silicone locking liners for suspension. A digital level gauge (DP200Hi, STS Co. Ltd., Japan) was used to check the angular changes of each condition.

An experimental prosthesis with the instrumented prosthetic pyramid (Europa).

The instrumented prosthetic pyramid was installed at the bottom of each participant's prosthetic socket. Baseline alignment was set up by a prosthetist based on observation, participants’ preferences, and the Compas system (Orthocare Innovations LLC, Edmonds,WA, USA)30. Alignment in the sagittal plane included the following conditions: (1) three-degree flexion/extension of the socket, anterior/posterior translation of the socket, (2) three-degree dorsiflexion/plantarflexion of the foot. Alignment conditions in the coronal plane included: (3) six-degree adduction/abduction of the socket, medial/lateral translation of the socket, (4) six-degree inversion/eversion of the foot. Adjustment in the sagittal plane was limited in order to compensate for the heel height of the prosthetic foot33. These conditions were set with respect to each individual's baseline alignment. The translations were established with equal displacement with angular changes33 as shown in Figs. 1 and 3. The mean (standard deviation) of translational perturbations were 22.67 (3.49) mm in six-degree angular changes and 11.35 (1.75) mm in three-degree angular changes.

Comparison of alignment changes.

The participants were asked to walk on a 15-m walkway at a self-selected speed, and data were collected under each alignment condition. The intervention was one-side blinded: the conditions of alignment were set by the same prosthetist in randomized orders and the participants were not informed of the order of alignment conditions. All the participants were accustomed to walking without footwear due to their local culture, therefore shoes were not used and their potential influence on gait was eliminated. Participants were given sufficient time to acclimatize themselves to the 15-m walkway before each condition.

To collect and export data of the socket reaction moments for data processing, Compas version 1.3.2 (Orthocare Innovations LLC, Edmonds,WA, USA) was used. In the sagittal plane, maximum flexion moments, % stance of maximum flexion moments, maximum extension moment, % stance of maximum extension moment, and zero-cross (the timing when the moments cross zero) were measured and averaged. In the coronal plane, moments of 5%, 20%, and 75% stance were measured and averaged. Negative values of coronal/sagittal socket reaction moments were defined as varus/flexion moments, respectively. The magnitudes of moments were normalized by participants’ body mass. These parameters were selected according to previous studies23,29.

The statistical analysis was performed in four comparison groups (Fig. 3) as follows:

Comparison #1: the effects of the socket alignment changes in the sagittal plane on the coronal socket reaction moments, including baseline-flexion-posterior translation and baseline-extension-anterior translation.

Comparison #2: the effects of the foot alignment changes in the sagittal plane on the coronal socket reaction moments, including baseline-plantarflexion-dorsiflexion.

Comparison #3: the effects of the socket alignment changes in the coronal plane on the sagittal socket reaction moments, including baseline-adduction-medial translation and baseline-abduction-lateral translation.

Comparison #4: the effects of the foot alignment changes in the coronal plane on the sagittal socket reaction moments, including baseline-inversion-eversion in the coronal plane.

Walking speeds in each alignment condition were also averaged and compared within these comparison groups.

When normal distribution was confirmed by Shapiro–Wilk tests, repeated measures analysis of variance (ANOVA) was performed, and otherwise, Friedman tests were performed, which was followed by post hoc tests with the Bonferroni method for multiple comparisons. Effect size was also calculated for each comparison (Partial eta squared for repeated measures ANOVA / Kendall's W for Friedman tests). P < 0.05 was defined as significant and SPSS ver. 26 (IBM Corporation, USA) was used for the statistical analysis.

According to the results of Shapiro–Wilk tests, repeated measures ANOVA were performed for all comparisons except for 20% stance of coronal socket reaction moment among baseline, posterior translation and flexion, and zero-cross and percent stance of peak extension moment among baseline, adduction, and medial translation, for which Friedman tests were performed.

There were no significant main effects on the walking speeds (P = 0.66: comparisons among baseline, anterior translation, and extension, and P = 0.90: comparisons among baseline, posterior translation, and flexion) (Table 2). There was a significant main effect at 20% stance among the baseline-posterior translation-flexion (P = 0.02). Post hoc tests indicated significant differences between the posterior translation and baseline (P = 0.007), as well as the flexion and baseline (P = 0.044, Table 3, Fig. 4A, B).

Socket reaction moment in the coronal plane under sagittal socket alignment changes (Comparison #1). (A) Anterior translation versus extension of the socket. Small line graphs indicate percent stance of socket reaction moments (5, 20, and 75%). (B) Posterior translation versus flexion of the socket. Small line graphs indicate percent stance of socket reaction moments (5, 20, and 75%). Whiskers indicate standard deviations. An asterisk (*) indicates P < 0.05. A dagger (†) indicates P < 0.01. Abbreviations: BL: baseline, AT: anterior translation, EX: extension, PT: posterior translation, FL: flexion, %stance: percent stance. N.S.: not significant.

There were no significant main effects on walking speed (P = 0.67: comparisons among baseline, plantarflexion, and dorsiflexion) (Table 2). There were no significant main effects among the conditions in parameters related to magnitudes and timings of socket reaction moments (Table 3, Fig. 5A).

Socket reaction moment under foot alignment changes. (A) Socket reaction moment in the coronal plane under sagittal alignment changes of prosthetic feet (dorsiflexion vs. plantarflexion) (Comparison #2). Small line graphs indicate 5, 20, and 75 percent stance of socket reaction moments. Whiskers indicate standard deviations. (B) Socket reaction moment in the sagittal plane under coronal alignment changes of prosthetic feet (inversion vs. eversion) (Comparison #4). Small line graphs indicate peak flexion and extension socket reaction moments, percent stance of peak flexion and extension socket reaction moments and zero-cross. Whiskers indicate standard deviations. Abbreviations: BL: baseline, DF: dorsiflexion, PF: plantarflexion, IV: inversion, EV: eversion, N.S: not significant.

There were no significant main effects on walking speed (P = 0.46: comparisons among baseline, lateral translation, and abduction, and P = 0.23: comparisons among baseline, medial translation, and adduction) (Table 2). There were no significant main effects among the conditions in parameters related to magnitudes and timings of socket reaction moments (Table 4, Fig. 6A, B).

Socket reaction moment in the sagittal plane under coronal socket alignment changes (Comparison #3). (A) Abduction versus lateral translation of the socket. Small line graphs indicate peak flexion and extension socket reaction moments, Percent stance of peak flexion and extension socket reaction moments and zero-cross. (B) Adduction versus medial translation of the socket. Small line graphs indicate peak flexion and extension socket reaction moments, percent stance of peak flexion and extension socket reaction moments, and zero-cross. Whiskers indicate standard deviations. Abbreviations: BL: baseline, AB: abduction, AD: adduction, LT: lateral translation, MT: medial translation, %stance: percent stance, N.S.: not significant.

There were no significant main effects on walking speed (P = 0.76: comparisons among baseline, inversion, and eversion) (Table 2). There were no significant main effects among the conditions in parameters related to magnitudes and timings of socket reaction moments (Table 5, Fig. 5B).

This study investigated the effects of alignment changes of the transtibial prosthetic socket (i.e., flexion/extension, abduction/adduction, and anterior/posterior, medial/lateral translation with equal displacement to counterpart angular alignment) and the foot (i.e., plantarflexion/dorsiflexion, inversion/eversion) on the out-of-plane socket reaction moments in order to determine the appropriate sequence of dynamic alignment. As walking speeds were not significantly different within each comparison group, alignment changes should be the only factor contributing to changes in the socket reaction moments.

The results showed that only the alignment changes of the socket in the sagittal plane (i.e., posterior translation and flexion) significantly increased the varus moment in the coronal plane. This was consistent with the previous study, which found that the extension of the socket significantly decreased varus moment in the coronal plane32. The alignment changes of the foot in the coronal and sagittal planes did not show significant changes in socket reaction moments in the out-of-planes (i.e., effects of sagittal alignment changes on coronal socket reaction moment, and effects of coronal alignment changes on sagittal socket reaction moment). This result was also in line with the previous study32. A prior study found that the alignment changes of the prosthetic foot in the transverse plane (i.e., toe-in and toe-out) significantly affected the coronal socket reaction moment29. This was the only significant effect of foot alignment changes on the out-of-plane moment.

The sagittal alignment of the socket affects the magnitude of flexion moment in the early stance when the ground reaction force is anterior to the knee joint37, and this may be a result of braking force originated from the residual limb and transferred to the socket to control knee extension. In the late stance, the alignment affects the extension moment to control knee flexion, as the ground reaction force vector during propulsion is directed behind the knee joint. When the transition of these forces are interrupted, the prosthesis may not move forward smoothly, leading to abnormal stress at the residuum-socket interface that likely affects the coronal moment. In contrast, the coronal alignment changes may not affect the sagittal socket reaction moment because six degrees of alignment changes might not be enough to influence braking and propulsion force. In clinical practice, angular adjustments greater than six degrees from the baseline alignment may not be adopted because it approaches the limit of angular changes for the components designed for daily use. Further study is needed to investigate the relationship between the sagittal alignment and coronal moment in a systematic and comprehensive way.

Previous studies revealed systematic effects of sagittal alignment changes of both the socket and the foot on sagittal socket reaction moments, and the effects of coronal alignment changes of the socket and the foot on coronal socket reaction moments27,38,39. It is likely that the sagittal socket reaction moment is affected merely by sagittal alignment changes of both the socket and the foot, whereas the coronal socket reaction moment is affected by alignment changes in all three planes: coronal alignment changes of both the socket and the foot, sagittal alignment changes of the socket, and transverse alignment changes of the foot. Therefore, it is suggested to first determine the sagittal alignment, because even if coronal or transverse alignment changes are performed after sagittal alignment is completed, these changes may not significantly affect the sagittal socket reaction moments. Although sagittal alignment changes may potentially influence the coronal socket reaction moment, the coronal moment can be tweaked by alignment changes in other planes (i.e., coronal and transverse) during dynamic alignment.

To establish coronal alignment, it might be necessary to consider the effects of transverse alignment. It is recommended that toe-in and toe-out angles should be determined according to the intact side at static alignment12. This could be mainly due to the aesthetic preference of the users. Therefore, toe-in/toe-out angles are generally determined at static alignment. However, the effect of transverse alignment determined during static alignment should also be considered in dynamic alignment because the transverse alignment changes systematically affect the coronal socket reaction moment29. For example, when the prosthetic foot is internally rotated (i.e., toe-in) at dynamic alignment, it may increase the magnitude of varus moment in the late stance during gait when compared to bench alignment29. Greater varus moment can also be induced by excessive coronal alignment changes (e.g., eversion of the foot39 or lateral translation/abduction of the socket23,27,38). Thus, in this case, there should be three options for alignment adjustments in transverse and/or coronal planes (i.e., decrease in toe-in angle, decrease in eversion, and decrease in lateral translation) and further adjustment of sagittal alignment may not be necessary.

There was no significant difference related to the timing of the socket reaction moment in the out-of-plane (i.e., effects of sagittal alignment changes on the timing of the coronal socket reaction moment, and effects of coronal alignment changes on the timing of the sagittal socket reaction moment). Our previous study suggested that in the sagittal plane angular changes mainly affect the timing of the socket reaction moment and translational changes affect primarily the magnitude of the socket reaction moment33. Also, this study suggested that translational changes in the coronal plane affect the magnitude of socket reaction moment primarily in the mid-to-late stance and angular changes show a similar effect, but to a lesser degree in the late stance29. Furthermore, it should be noted that the effects of angular changes in the socket are equivalent to the combined effects of angular and translational changes in the foot (Fig. 1)33,39. Therefore, whether an angular or translational adjustment should be performed first should be determined based on their effect on timing and/or magnitude of socket reaction moment.

This study has some limitations. As male participants were predominant, it is hard to further elucidate the possible effect of gender. As the length of residual limb and leg length varied among participants, the same amount of angular change could introduce different displacements (d in Fig. 1) at the distal end. Participants in the study have relatively high activity levels (K3-4) and it remains unclear if similar findings could be revealed in populations with low activity levels (Table 1). The prosthetic foot was controlled in this study and the effects of different prosthetic foot design remain unclear. The acclimation period was also brief so the long-term effects are unknown.

In conclusion, this study showed that sagittal alignment may affect both the sagittal and coronal socket reaction moments, and coronal alignment may affect only the coronal socket reaction moment. According to these findings, the appropriate sequence of adjusting the alignment of transtibial prostheses should be first in the transverse plane in static alignment (i.e., adjustment of toe-in/toe-out angles for standing symmetry)29, followed by dynamic alignment (i.e., adjustment of toe-in/toe-out angles for gait symmetry), then in the sagittal alignment in dynamic alignment procedure, and lastly in the coronal plane (Fig. 7). It should be noted that the coronal socket reaction moment may be affected by alignment changes in all three planes. Therefore, our hypothesis is supported, and the adjustment of coronal alignment should be performed lastly considering the out-of-plane effects. These findings could serve as a clinical guideline for alignment processes in transtibial prostheses. However, future studies with a larger sample size, more representative population with a wider selection of prosthetic feet and socket designs is needed to validate this guideline and implementation in clinical practice.

A flowchart of dynamic alignment using the socket reaction moment. Abbreviation: SRM: socket reaction moment.

The datasets generated for this study are available from the corresponding author on reasonable request.

Childress, W. L. & Wurdeman, S. S. R. Transtibial amputation: prosthetic management. In Atlas of Amputations and Limb Deficiencies (eds Krajbich, J. I. et al.) 493–508 (American Academy of Orthopaedic Surgeons, 2016).

Google Scholar

Hannah, R. E., Morrison, J. B. & Chapman, A. E. Prostheses alignment: Effect on gait of persons with below-knee amputations. Arch. Phys. Med. Rehabil. 65, 159–162 (1984).

CAS Google Scholar

Pinzur, M. S. et al. The effect of prosthetic alignment on relative limb loading in persons with trans-tibial amputation: A preliminary report. J. Rehabil. Res. Dev. 32, 373–377 (1995).

CAS Google Scholar

Rossi, S. A., Doyle, W. & Skinner, H. B. Gait initiation of persons with below-knee amputation: The characterization and comparison of force profiles. J. Rehabil. Res. Dev. 32, 120–127 (1995).

CAS Google Scholar

Schmalz, T., Blumentritt, S. & Jarasch, R. Energy expenditure and biomechanical characteristics of lower limb amputee gait. Gait Posture 16, 255–263 (2002).

Article Google Scholar

Fridman, A., Ona, I. & Isakov, E. The influence of prosthetic foot alignment on trans-tibial amputee gait. Prosthet. Orthot. Int. 27, 17–22 (2003).

Article CAS Google Scholar

Chow, D. H. K., Holmes, A. D., Lee, C. K. L. & Sin, S. W. The effect of prosthesis alignment on the symmetry of gait in subjects with unilateral transtibial amputation. Prosthet. Orthot. Int. 30, 114–128 (2006).

Article Google Scholar

Frossard, L., Beck, J., Dillon, M. & Evans, J. Development and preliminary testing of a device for the direct measurement of forces and moments in the prosthetic limb of transfemoral amputees during activities of daily living. JPO J. Prosthet. Orthot. 15, 135–142 (2003).

Article Google Scholar

Grumillier, C., Martinet, N., Paysant, J., André, J. M. & Beyaert, C. Compensatory mechanism involving the hip joint of the intact limb during gait in unilateral trans-tibial amputees. J. Biomech. 41, 2926–2931 (2008).

Article CAS Google Scholar

Neumann, E. S., Brink, J., Yalamanchili, K. & Lee, J. S. Regression estimates of pressure on transtibial residual limbs using load cell measurements of the forces and moments occurring at the base of the socket. J. Prosthet. Orthot. 25, 1–12 (2013).

Article Google Scholar

Fiedler, G. & Johnson, M. S. Correlation of transtibial prosthetic alignment quality and step-by-step variance of gait. J. Prosthet. Orthot. 29, 19–25 (2017).

Article Google Scholar

New York University Post-Graduate Medical School Prosthetics and Orthotics. Gait analysis of below-knee amputee. in Lower-Extremity Prosthetics 171–176 (New York University, 1971).

Radcliffe, C. W. & Foort, J. The Patellar-Tendon Bearing Below-Knee Prosthesis (University of California Press, 1961).

Google Scholar

Ikeda, A. J. et al. Á priori alignment of transtibial prostheses: A comparison and evaluation of three methods. Disabil. Rehabil. Assist. Technol. 7, 381–388 (2012).

Article Google Scholar

Wu, Y., Brncick, M. D., Krick, H. J. & Scotchcast, P. V. C. Interim prosthesis for below-knee amputees. Bull. Prosthet. Res. 18, 40–50 (1981).

Google Scholar

Reisinger, K. & Casanova, H. Comparison of á priori alignment techniques for transtibial prostheses in the developing world-pilot study. Disabil. Rehabil. 29, 863–872 (2007).

Article Google Scholar

Reisinger, K. D., Casanova, H., Wu, Y. & Moorer, C. Evaluation and comparison of á priori alignment techniques for transtibial prostheses in the developing world—field trial in Nicaragua. Disabil. Rehabil. Assist. Technol. 4, 393–405 (2009).

Article Google Scholar

Isakov, E., Mizrahi, J., Susak, Z., Ona, I. & Hakim, N. Influence of prosthesis alignment on the standing balance of below-knee amputees. Clin. Biomech. 9, 258–262 (1994).

Article CAS Google Scholar

Blumentritt, S., Schmalz, T., Jarasch, R. & Schneider, M. Effects of sagittal plane prosthetic alignment on standing trans-tibial amputee knee loads. Prosthet. Orthot. Int. 23, 231–238 (1999).

Article CAS Google Scholar

Mooney, R. et al. Radiographic parameters improve lower extremity prosthetic alignment. J. Child Orthop. 7, 543–550 (2013).

Article Google Scholar

Brinkmann, J. & Stevens, P. Clinical consideration of observational gait analysis. In Atlas of Amputations and Limb Deficiencies (eds Krajbich, J. et al.) 81–95 (American Academy of Orthopaedic Surgeons, 2016).

Google Scholar

Boone, D. A. et al. Perception of socket alignment perturbations in amputees with transtibial prostheses. J. Rehabil. Res. Dev. 49, 843 (2012).

Article Google Scholar

Hashimoto, H., Kobayashi, T., Gao, F., Kataoka, M. & Okuda, K. The effect of coronal prosthetic alignment changes on socket reaction moments, spatiotemporal parameters, and perception of alignment during gait in individuals with transtibial amputation. J. Rehabil. Assist. Technol. Eng. 5, 205566831879540 (2018).

Google Scholar

Beyaert, C., Grumillier, C., Martinet, N., Paysant, J. & André, J.-M. Compensatory mechanism involving the knee joint of the intact limb during gait in unilateral below-knee amputees. Gait Posture 28, 278–284 (2008).

Article CAS Google Scholar

Jonkergouw, N., Prins, M. R., Buis, A. W. P. & van der Wurff, P. The effect of alignment changes on unilateral transtibial amputee's gait: A systematic review. PLoS ONE 11, e0167466 (2016).

Article Google Scholar

Zahedi, M. S., Spence, W. D., Solomonidis, S. E. & Paul, J. P. Alignment of lower-limb prostheses. J. Rehabil. Res. Dev. 23, 2–19 (1986).

CAS Google Scholar

Boone, D. A. et al. Influence of malalignment on socket reaction moments during gait in amputees with transtibial prostheses. Gait Posture 37, 620–626 (2013).

Article Google Scholar

Kobayashi, T., Orendurff, M. S. & Boone, D. A. Dynamic alignment of transtibial prostheses through visualization of socket reaction moments. Prosthet. Orthot. Int. 39, 512–516 (2015).

Article Google Scholar

Hashimoto, H. et al. The effect of transverse prosthetic alignment changes on socket reaction moments during gait in individuals with transtibial amputation. Gait Posture 65, 8–14 (2018).

Article Google Scholar

Chen, C. W. J. et al. Evaluation of an instrument-assisted dynamic prosthetic alignment technique for individuals with transtibial amputation. Prosthet. Orthot. Int. 40, 475–483 (2016).

Article Google Scholar

Staros, A. Dynamic alignment of artificial legs with the adjustable coupling. Artif. Limbs 7, 31–43 (1963).

CAS Google Scholar

Kobayashi, T., Orendurff, M. S., Zhang, M. & Boone, D. A. Effect of transtibial prosthesis alignment changes on out-of-plane socket reaction moments during walking in amputees. J. Biomech. 45, 2603–2609 (2012).

Article Google Scholar

Hashimoto, H., Kobayashi, T., Kataoka, M. & Okuda, K. Angulation vs translation of transtibial prosthetic socket: Their difference analyzed by socket reaction moments. Gait Posture 97, 137–146 (2022).

Article Google Scholar

Kobayashi, T., Orendurff, M. S. & Boone, D. A. Effect of alignment changes on socket reaction moments during gait in transfemoral and knee-disarticulation prostheses: Case series. J. Biomech. 46, 2539–2545 (2013).

Article Google Scholar

Kobayashi, T. et al. Effect of alignment changes on socket reaction moments while walking in transtibial prostheses with energy storage and return feet. Clin. Biomech. 29, 47–56 (2014).

Article Google Scholar

Plug-in Gait Reference Guide. Vicon Motion Systems Limited. https://docs.vicon.com/display/Nexus210/PDF+downloads+for+Vicon+Nexus?preview=/98963539/98964068/Plug-inGait ReferenceGuide.pdf. 2020 (2020).

Perry, J. & Burnfield, J. M. Knee. in Gait Analysis 85–102 (Slack, 2010).

Kobayashi, T., Orendurff, M. S., Arabian, A. K., Rosenbaum-Chou, T. G. & Boone, D. A. Effect of prosthetic alignment changes on socket reaction moment impulse during walking in transtibial amputees. J. Biomech. 47, 1315–1323 (2014).

Article Google Scholar

Hashimoto, H., Kobayashi, T., Kataoka, M. & Okuda, K. Influence of coronal and sagittal prosthetic foot alignment on socket reaction moments in transtibial prostheses during walking. Gait Posture 90, 252–260 (2021).

Article Google Scholar

Download references

The authors would like to thank all individuals who participated in this study. The authors have received no financial support.

Osaka Metropolitan University, Habikino City, Osaka, Japan

Hiroshi Hashimoto & Masataka Kataoka

Pacific Supply Co. Ltd., Daito City, Osaka, Japan

Hiroshi Hashimoto

Department of Biomedical Engineering, Faculty of Engineering, The Hong Kong Polytechnic University, Hung Hom, Kowloon, Hong Kong, China

Toshiki Kobayashi

Department of Kinesiology and Health Promotion, University of Kentucky, Lexington, KY, USA

Fan Gao

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

You can also search for this author in PubMed Google Scholar

This project was conceived by H.H. and T.K. Data collection were conducted by H.H. Data analyses were conducted by H.H. and T.K. The manuscript was drafted by H.H. and T.K. Critical review and edits of the manuscript was done by F.G. and M.K. T.K. and M.K. supervised the study. All authors read and approved the final manuscript.

Correspondence to Toshiki Kobayashi.

HH is an employee of Pacific Supply Co. Ltd., the distributor of Europa in Japan. TK was an employee of Orthocare Innovation LLC, the manufacturer of Europa. FG and MT declare no potential conflict of interest.

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/.

Reprints and Permissions

Hashimoto, H., Kobayashi, T., Gao, F. et al. A proper sequence of dynamic alignment in transtibial prosthesis: insight through socket reaction moments. Sci Rep 13, 458 (2023). https://doi.org/10.1038/s41598-023-27438-1

Download citation

Received: 22 October 2022

Accepted: 02 January 2023

Published: 10 January 2023

DOI: https://doi.org/10.1038/s41598-023-27438-1

Anyone you share the following link with will be able to read this content:

Sorry, a shareable link is not currently available for this article.

Provided by the Springer Nature SharedIt content-sharing initiative

By submitting a comment you agree to abide by our Terms and Community Guidelines. If you find something abusive or that does not comply with our terms or guidelines please flag it as inappropriate.